← Back to CVE List
CVE-2026-7251NVD
Vulnerability Summary
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS v4.0 Base Metrics — Score 9.3 (CRITICAL)
Attack VectorNetwork
Attack ComplexityLow
Attack RequirementsNone
Privileges RequiredNone
User InteractionNone
Confidentiality (Vulnerable System)High
Integrity (Vulnerable System)High
Availability (Vulnerable System)High
Confidentiality (Subsequent System)None
Integrity (Subsequent System)None
Availability (Subsequent System)None
CVSS v3.1 Base Metrics — Score 9.8 (CRITICAL)
Attack VectorNetwork
Attack ComplexityLow
Privileges RequiredNone
User InteractionNone
ScopeUnchanged
ConfidentialityHigh
IntegrityHigh
AvailabilityHigh
Affected & Patched Versions
- Eppendorf BioFlo 320 >= All
Not provided by cveorg for this CVE.